• Harvests autologous epithelium
• No donor site anesthesia required
• Automated, precise, and reproducible epidermal harvesting
• Minimal pain and donor-site trauma
The 3M™ Cellutome™ Epidermal Harvesting System is intended to reproducibly cut a thin skin graft for autologous skin grafting.
The harvester is a disposable component and is intended for single use only. Reuse of disposable components may result in contamination and infection.
Inspect the 3M™ Cellutome™ Epidermal Harvesting System before use for signs of visible damage. Do not use if there are any indications of damage.
Harvest from areas of healthy intact skin (donor site).
Actuating the harvester handle prior to complete microdome formation may result in patient pain and/or bleeding.
Actuating the harvester handle prior to placement of the dressing may result in an inability to secure the micrografts for application.
Wipe the vacuum head and the control unit nest following institutional procedures used for cleaning and disinfection of other hard surface durable electronic medical equipment between each patient use. Inadequate cleaning may result in patient contamination.
Use the harvester only with the 3M™ Cellutome™ Epidermal Harvesting System control unit and vacuum head.
Discard all disposable items (harvester and dressing) in accordance with the local medical waste disposal regulations.
Handle the vacuum head with care. Do not drop the vacuum head, as doing so could damage the glass warming window. Do not suspend or hold the vacuum head by the tubing.
Reversible maceration at the recipient site is a potential risk. Please use best clinical judgement for preventing maceration.
Temperature Range: 0°F (-18°C) to 140°F (60°C)
Relative Humidity Range: 15%-95% non-condensing
Automated | ● |
Bedside procedure | ● |
Dressings included | |
Includes epidermal and dermal layers | |
Includes epidermal layers | ● |
Local anesthesia required | |
No anesthesia required | ● |
Single use | ● |
Educational Material Available | ● |
Free Samples/Trials Available | ● |
Published Clinical Article Available |
Chronic Wounds
Granulating/Epithelializing Wounds
Pressure Ulcers
Superficial Wounds
Venous Ulcers
3M™ Adaptic™ Touch Non-Adhering Silicone Dressing (for transferring harvested epidermal cells to the wound site).
Bolster dressing of choice.
12 month, 3 year harvester shelf life.
Harvester: 5.7"x3.7"x6", 0.2lb
Vacuum head: 5.5"x3.9"x1.3", 0.6lb
Control unit: 10.6"x9.3"x9.4", 7.3lb
Control unit basket accessory: 2lb maximum weight rating.
Electrical:
Internal power supply input: 100-240 VAC 3.2A 50/60Hz
Internal power supply output: 24 VDC, 10.4A
Power inlet fuse:
Type: 5x20mm time lag glass cartridge
Current rating: 5A
Voltage rating: 250V
Operative speed(s): 600ms
3M focuses on providing better care through patient-centered science. From wound and skin care to solutions for BSI and SSI risk reduction, 3M is ready to partner with health care providers to help transform outcomes.